Checking the calibration report
The calibration report must be checked and saved for QC purposes. To obtain the report, type $B,1,CR or click on the Calibration Report in the Raw Communication window.
This is a report consistent with a good calibration:
Check for compliance with the following QC standards:
Parameter | QC Limits | Comment | |||||
S/N | Must be a valid S/N | ||||||
Model |
#91 for SGEO2 devices #102 for ST-MOD Rev B #103 for ST-MOD Rev C |
Model numbers may vary. Check with your supervisor if unsure. | |||||
RTC Single-Point (clock error) | -50000 to 50000 ppb | Larger errors may indicate a clock circuit problem | |||||
RTC Single-Point (temperature) | 12000 – 22000 mdeg C | Clock calibration temperature within +/- 5 degrees of the average world ocean surface temperature of 17 deg. C | |||||
Magnetometer Voltage Response Calibration |
|
||||||
Magnetometer Gain-Offset Calibration | A numerical result for axis X,Y, and Z. | ||||||
|
0-29 ppt (parts per thousand) |
A higher value indicates an imprecise gain-offset calibration. Repeat gain-offset calibration and verification
|